Products Category
+ pH Electrode
Details
Applications
Features
Product #: 368417,00000
Applications
-
Industrial and waste water treatment applications
Features
- Double junction with gelified electrolyte KCl
- Sealed reference design Ag/AgCl
- Withstands high temperature up to 110°C
- Not contaminated by sulfide or polluting ions
- Operates in process under pressure
Cable |
|||
Model
|
Reference
|
Length
|
Max temperature
|
AS9 110
|
359016,10110
|
3 m/10 ft
|
70°C/158°F
|
AS9 120
|
359016,10120
|
10 m/33 ft
|
70°C/158°F
|
AS9 122
|
359016,10122
|
20 m/66 ft
|
70°C/158°F
|
AS9 HT 130
|
359016,10130
|
3 m/10 ft
|
110°C/230°F
|
AS9 HT 160
|
359016,10160
|
1.5 m/5 ft
|
110°C/230°F
|
The cables have double shields in order to conform to European directive 89/336/CEE
regarding electromagnetical compatibility (EN 50081 & EN 50082). External shields are
not linked to the connector side : they have to be connected on the transmitter side.
regarding electromagnetical compatibility (EN 50081 & EN 50082). External shields are
not linked to the connector side : they have to be connected on the transmitter side.
Specifications
- Part number 368417,00000 Dimensions 12mm diameter x 120mm length Measuring range 0-14 pH Temperature -5 to 110°C Max pressure 10 bar at 25°C Stability 0.02 pH over 24 hours (T, P constant) Response time <15s at 25°C Glass impedance 100 MΩ Probe material Glass membrane , ceramic junction Probe mounting Standard PG13.5 threaded cap