+ pH meter
POLYMETRON 9123 pH Calculator
  
Details

PH Meter  polymetron 9123 pH Calculator

09123=A=7000
09123=A=7010

   Product #: 09123=A=0011
    Dual conductivity/pH calculated transmitter Polymetron 9123

General
    The transmitter shall measure conductivity on
    two channels with selectable units for
    conductivity and/or resistivity.

   The transmitter shall deliver the pH through
   the measurement of two conductivity probes
   placed before and after an external cationic
   resin bed.

   The transmitter menu should allow for the
   choice of the water chemistry being used:
   NaOH, NH3 or ethanolamine

Enclosure
   The 9123 transmitter shall have a NEMA4X,
   IP65 certified enclosure in aluminum and be
   supplied with a universal mounting kit for
   surface, panel and pipe mounting.
   The transmitter shall be ISO 9001-
   2000manufactured, be provided with a
   certificate of calibration and comply with CE
   and UL compliances.
Display
   The analyzer shall have a graphical dot
   matrix LCD display with 80 x 64 pixels and
   LED backlighting. The main display shall
   be 34 x 67.4 mm (1.3 x 2.7 in) with 5 lines
   of 13 characters.
Display information
   The display shall show the calculated pH
   alone or combined with conductivity of
   channel 1and channel 2.

   Simultaneously main displayed information
   shall contain measure, temperature and 4
   relay alarms status. Auxiliary displayed
   information shall be available without
   measurement interruption through a “one
   button” operation and provide the complete
   view of the analyzer status (alarm
   assignments, analog output setup and
   parameters, electrodes polarization
   information).
User interface
   The analyzer shall have screens for
   transmitter operation in six languages
   (English, French, German, Spanish, Dutch
   and Italian).
Calibration modes
   The 9123 shall have two main calibration
   methods: Electric in two point with a reference
   resistor and air, 1 point with external standard
   (slope only) and 2 point with external standard
   + air.

Calibration data
   The analyzer shall have a calibration logbook,
   recording data of the last two calibrations
   (date, value, zero).

Security
   The analyzer shall have three password
   protected access levels for transmitter
   calibration, programming and servicing.

Alarms
   The analyzer shall have 4 alarm relays
   and settings for 3 main parameters:
 
 Service Parts
 
09123=A=0000
9123 pH calc transmitter
09123=A=0004
9123 pH calc transmitter + 4-relay board
09123=A=0011
9123 pH calc transmitter + RS485
09123=A=0012
9123 pH calc transmitter + Profibus DP
09123=A=0015
9123 pH calc transmitter + 4-relay and RS485 boards
09123=A=0016
9123 pH calc transmitter + 4-relay and Profibus DP
09123=A=0020
9123 pH calc transmitter(24 V)
09123=A=0024
9123 pH calc transmitter + 4-relay board (24V)
09123=A=0031
9123 pH calc transmitter + RS485 board (24 V)
09123=A=0032
9123 pH calc transmitter + Profibus DP (24 V)
09123=A=0035
9123 pH calc transmitter + 4-relay + RS485 (24 V)
09123=A=0036
9123 pH calc transmitter + 4-relay board and Profibus DP (24 V)
09123=A=1000
9123 pH calc analyzer
09123=A=1004
9123 pH calc analyzer + 4xrelay board
09123=A=1011
9123 pH calc analyzer + RS485
09123=A=1012
9123 pH calc analyzer + Profibus DP
09123=A=1015
9123 pH calc analyzer + 4xrelay and RS485 boards
09123=A=1016
9123 pH calc analyzer + 4xrelay and Profibus DP
09123=A=1020
9123 pH calc analyzer (24 V)
09123=A=1024
9123 pH calc analyzer + 4xrelay board (24V)
09123=A=1031
9123 pH calc analyzer + RS485 board (24 V)
09123=A=1032
9123 pH calc analyzer + Profibus DP (24 V)
09123=A=1035
9123 pH calc analyzer + 4xrelay + RS485 (24 V)
09123=A=1036
9123 pH calc analyzer + 4xrelay board and Profibus DP (24 V)
09123=A=3000
9123 pH calc transmitter tropicalised
09123=A=6200
Replacement kit with a 91XX CPU board programmed with dual conductivity firmware and mounting instructions.
09123=A=7000
Spare resin cartridge complete with resin
09123=A=7010
Resin kit (includes 2 filters, 2L sachet of resin, funnel, instructions)
09123=A=8001
Cable for 9123 (1 meter) with IP 65 connector
09123=C=0100
Dual measurement cell
09123=C=0200
Cationic resin cartridge (empty)
09123=C=0300
Resin cartridge closure
09123=C=0400
Inner tube for resin cartridge, ss316
09123=C=0500
Flat gasket, Viton
09123=C=0600
Resin cartridge filter
09123=T=0000
9123 transmitter test certificate